AirCurve 10 VAuto: Clinical Menu


HOME
DIAL
To access the Clinical Menu:
- 
Press and hold the Dial and Home buttons simultaneously for 3 seconds.
 
MODE
- 
CPAP - treatment at fixed pressure.
 
SET PRESSURE
Fixed treatment pressure in CPAP mode:
- 
4.0 - 20
 
- 
S - treatment at fixed inspiratory and expiratory pressures.
 
IPAP
Inspiratory pressure:
- 
4.0 - 25
 
EPAP
Expiratory pressure:
- 
3.0 - 25
 
EASY-BREATHE
Mimics natural breathing pattern and reduces the sound of the device.
- 
Off
 - 
On
 
RISE TIME*
Time for pressure to increase from EPAP to IPAP (*enabled with Easy-Breathe Off).
- 
Min, 150 - 900ms
 
TI MAX
Maximum inspiratory time:
- 
0.3 - 4s
 
TI MIN
Minimum inspiratory time:
- 
0.1 - 4s or Ti Max
 
TRIGGER
Sensitivity level which the devices changes from EPAP to IPAP:
- 
Very Low
 - 
Low
 - 
Medium
 - 
High
 - 
Very High
 
CYCLE
Sensitivity level which the devices changes from IPAP to EPAP:
- 
Very Low
 - 
Low
 - 
Medium
 - 
High
 - 
Very High
 
- 
VAuto - treatment at auto-adjusting Inspiratory and expiratory pressures.
 
MAX IPAP
MIN EPAP
PS
Maximum inspiratory pressure:
- 
4.0 - 25
 
Minimum expiratory pressure:
- 
4.0 - 25
 
Pressure support (difference between inspiratory and expiratory pressures):
- 
4.0 - 10
 
TI MAX
Maximum inspiratory time:
- 
0.3 - 4s
 
TI MIN
Minimum inspiratory time:
- 
0.1 - 4s or Ti Max
 
TRIGGER
Sensitivity level which the devices changes from EPAP to IPAP:
- 
Very Low
 - 
Low
 - 
Medium
 - 
High
 - 
Very High
 
CYCLE
Sensitivity level which the devices changes from IPAP to EPAP:
- 
Very Low
 - 
Low
 - 
Medium
 - 
High
 - 
Very High
 
MASK
Type of mask used during treatment:
- 
Pillows
 - 
Full Face
 - 
Nasal
 
RAMP TIME
Time period which the pressure increases from a lower starting pressure to the treatment pressure or min pressure:
- 
Off
 - 
5 - 45
 
START PRESSURE*
Pressure at the start of the ramp (*in CPAP mode):
- 
4 - 20 or set pressure
 
START EPAP*
Pressure at the start of the ramp (*in S and VAuto modes):
- 
4 - EPAP
 
EPR*
Expiratory Pressure Relief reduces mask pressure during exhalation for comfort (*in CPAP mode):
- 
Off
 - 
On: Ramp Only
 - 
On: Full Time
 
EPR LEVEL*
The amount of pressure that is reduced during exhalation (*in CPAP mode):
- 
1
 - 
2
 - 
3
 
HUMIDITY
The level the humidifier moistens the air you breathe during treatment:
- 
Off
 - 
1-8
 
CLIMATE CTRL*
Controls temperature and humidity levels during therapy (*enabled with ClimateLineAir heated tubing connected).
- 
Auto - automatically adjusts to maintain a constant humidity level of 85%.
 - 
Manual - you adjust Tube Temperature and Humidity Level separately.
 
TUBE TEMP*
The temperature inside the tube (*enabled with ClimateLineAir heated tubing connected):
- 
Auto
 - 
60–86°F
 - 
Off
 
TUBE*
Inner diameter of tubing used (*disabled with ClimateLineAir connected):
- 
SlimLine: 15mm
 - 
Standard: 19mm
 
AB FILTER
Select whether an antibacterial filter is attached:
- 
No
 - 
Yes
 
ESSENTIALS
Access level granted to patient at home screen:
- 
Plus: Advanced options and data access
 - 
On: Basic options and data access
 
LEAK ALERT
When enabled, leaks >40 L/min for >20 sec result in an audible alert and a high leak message is displayed.
- 
Off
 - 
On
 
SMARTSTART*
If enabled, unit will start automatically when patient breathes into the mask and stop automatically when mask is removed (*disabled with Leak Alert On):
- 
Off
 - 
On
 
REMINDERS
Unit will remind patients to replace mask and supplies every 0-24 months:
- 
Mask
 - 
Water Tub
 - 
Tube
 - 
Filter
 
LANGUAGE
- 
English
 - 
Francais
 - 
Espanol
 - 
Portugues
 
DATE
- 
Day / Month / Year
 
TIME
PRESS. UNITS
TEMP. UNITS
- 
Hours : Minutes
 
- 
cmH2O
 - 
hPA
 
- 
°C
 - 
°F
 
RESTORE DEFAULTS
- 
No
 - 
Yes
 
ERASE DATA
ABOUT
- 
No
 - 
Yes
 
Additional information about the unit such as Run Hours and internal modem (cellular provider, type and signal).